Soil Gas Monitoring with Flux Chamber

Flux_Chamber_White_Lab

Perform interstitial gas monitoring using a Flux Chamber to obtain a reliable measurement of volatile compounds at the soil-air interface.

The Flux Chamber used by Whitelab laboratories is of the open-dynamic type and allows both continuous sampling with field instruments and individual sampling with vials.

When hydrogeological conditions do not allow the direct measurement of soil gas with probes embedded in the soil, interstitial gas monitoring can be carried out using the Flux Chamber.

This method is particularly useful in environmental remediation contexts, where it can be used throughout the various stages of the process: from characterisation to final verification.

Monitoring using Flux Chamber leads to an estimate of the gas/vapour flux in terms of mass per unit area in the unit time (μg/m2*min).

The flux chamber isolates a portion of the area from external environmental conditions (wind, pollutants), enabling reliable measurement of volatile compounds at the soil-air interface.

The Flux Chamber used by Whitelab laboratories is of the open-dynamic type:
  • A stream of carrier gas (nitrogen 5.0) is injected into the flux chamber at a constant flow rate through a distribution pipe, ensuring homogeneous distribution
  • The chamber is placed on the ground and sealed by placing a layer of wet sand or dry bentonite around the base of the chamber to a height of approximately 5 cm
  • Prior to actual sampling a 'purge' of at least 4 volumes of the chamber is carried out to remove air and achieve conditions of perfect soil gas mixing within the chamber
    • During the purging phase, various parameters are recorded at regular intervals, such as: T° inside and outside the chamber
    • Concentrations of O2(%), CO2(%), CH4(%), COV (ppm)
    • ΔP (Pa) between inside and outside the chamber and atmospheric pressure
  • Sampling is performed using a perforated rod extending along the full height of the flow chamber, so as to sample the entire volume, and connecting it to one or more sampling lines through the outlet port.
The concentration of the substances sought in the soil gas can be measured:
  • Continuously, by means of analytical field instrumentation (PID - FID - GC/MS)
  • Through sampling with chemical desorption vials, thermal desorption vials, colorimetric vials, tedlar bags, canisters, etc.

Whitelab’s Flux Chamber has undergone performance testing in collaboration with the University of Bologna to ensure results of high scientific value.

Find out more about Whitelab Flux Chamber

Searchable substances

Flux Chamber searchable substances include:

  • Aliphatic, aromatic, halogenated and nitrogenous hydrocarbons
  • Sulphur compounds
  • Aldehydes and ketones
  • Mercury
  • Tetraethyl lead
  • Phenols
  • Cresols
  • Ammonia
  • Amines
  • PAHs
  • H2S
  • Pesticides
  • PCBs
Instrumental characteristics of the Flux Chamber
Type: Open Dynamic

Flat section: Ø 0.4 m

Internal volume, total chamber volume: 0.050 m3

Chamber height: 0.4 m

Base area: 0.126 m2

Insufflation system: circular tube closed at one end and perforated in several parts

Residence time (τ): 10 min

Material: stainless steel internally lined with PFA (Perfluoroalkoxy), a polymer inert to many chemical species, including mercury and aliphatic, aromatic, halogenated, nitrogenous hydrocarbons, etc. The perforated sampling rod is made of PTFE

Chamber: opaque

Carrier type: nitrogen 5.0

Carrier flow rate during purging: 5 L/min

Carrier flow rate during sampling: 5 L/min

Material for sealing to soil: Wet sand

Material and type of collector pipes: PTFE
Flux Chamber performance testing
Whitelab’s Flux Chamber is subject to the following performance checks carried out in collaboration with the University of Bologna:
  • Verification of the chamber blank for hydrocarbons C<12 and C>12: the concentrations of the analytes sought is lower than the limit values of quantification of the method used for sampling and testing

  • Verification of carrier gas saturation time inside the chamber: by which the purge time required before sampling was estimated

  • Verification of correct vertical diffusion of carrier gas inside the chamber: there is no stratification of carrier gas inside the flux chamber

  • Hydrocarbon and mercury recovery tests: the average recovery of individual compounds is 102%

  • Verification of the reproducibility of the estimate of the emissive flux from an artificial source: the variability of the data obtained by using the flux chamber is comparable with that obtained by the sampling and testing method alone

  • Verification of the correct distribution of a pollutant along the vertical axis of the flux chamber: there is no stratification of the pollutants inside the flux chamber and the sample that is taken through the sampling rod perforated at several points is representative of the average concentration of the pollutant inside the flux chamber

Would you like to receive more information or schedule a customised consultation for your company? Please, fill out the form and we will get back to you as soon as possible

"*" indicates required fields

Name and surname*
This field is for validation purposes and should be left unchanged.

Analysis and services

SERVICES

Legionella

Microbiological testing, sampling, risk assessment document, consultancy and Risk Assessment.

SERVICES

EC health mark

Testing and services for obtaining and maintaining the EC health mark.

SERVICES

Pharmaceutical packaging

Testing and services to test the pharmaceutical packaging effectiveness to ensure the maintenance of sterility.

SERVICES

Biocompatibility

Whitelab's offer for the biological evaluation of your Medical Device with biocompatibility studies.

SERVICES

REACH & RoHS

Testing and services to ensure the compliance of consumer products.